6 Commissioning
Topclean S
118
Hauser
Calibration
redox relative
"Data entry relative"
Entry of two % calibration points to which one mV value is assigned.
Calibration
redox relative
"Calibration relative"
For calibration, two vessels are filled with a sample of the medium. The contents of the first vessel
are detoxified and used as buffer 1.
The contents of the second vessel are left unchanged and used as buffer 2.
CHOICE
(default = bold)
INFO
Calibration with rel. data
entry
Note in display
Display of the type of on-site calibration selected in the calibration settings.
1. (0...30%):
1. Voltage
2. (70...100%)
2. Voltage
20 %
0600
mV
80 %
-0600
mV
Enter calibration points
In this field, create two measured value pairs (pair 1 and pair 2).
Measured value pair 1 in the range 0...30%: assign, for example, the voltage
0600 mV to the percentage value 20 %.
Measured value pair 2 in the range 70...100%: assign, for example, the voltage
–0600 mV to the percentage value 80 %.
The settings made become effective immediately after confirmation with
F
.
Offset too high / too low
Note in display:
Error message if the entered offset leaves the maximum range.
Accept
Cancel
Repeat calibration
End of calibration
Accept
: Pressing
F
accepts the new calibration data.
Cancel
: The data are not accepted, the calibration is not repeated.
Repeat calibration
: The data are rejected and a new calibration is started.
Electrode submersed?
Note in display:
Is the electrode back in the medium, so that measurement can take place?
CHOICE
(default = bold)
INFO
Calibration with rel.
calibration
Note in display
Display of the type of on-site calibration selected in the calibration settings.
Immerse electrode in buffer 1
Handling instructions
Immerse the electrode in buffer 1 / 2 (detoxified sample, see above).
Confirm by pressing
F
20 %
(0 ... 30 %)
Enter buffer
Enter the relative redox value of buffer 1 / 2 (detoxified sample) in percent.
Time: 10 s
mV 1:
225
Checks the stability of the calibration
Wait until the measurement is stable:
Time does not count down,
mV value no longer flashes,
Confirm by pressing
F
.
!
Note!
You can set the criteria for the stability check under "
Ñ
> Setup 1 > Calibration >
Cal. settings".
Invalid calibration value
Note in display:
Error message if the offset is too large.
Accept
Cancel
Repeat calibration
End of calibration
Accept
: Pressing
F
accepts the new calibration data.
Cancel
: The data are not accepted, the calibration is not repeated.
Repeat calibration
: The data are rejected and a new calibration is started.
Electrode submersed?
Note in display:
Is the electrode back in the medium, so that measurement can take place?
Summary of Contents for Topclean S CPC30
Page 154: ...10 Technical data Topclean S 154 Endress Hauser...
Page 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Page 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Page 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Page 159: ...Topclean S 11 Appendix Endress Hauser 159...
Page 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Page 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Page 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Page 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Page 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Page 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Page 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Page 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Page 176: ...Index Topclean S 176 Endress Hauser...
Page 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...