6 Commissioning
TopClean S CPC 30
88
EHauser
Calibration
redox relative
"Data entry relative"
Entry of two % calibration points to which one mV value is assigned.
CCB5
Offset
Good
0005 mV
Note in display:
Info on electrode.
Data on the offset and calibration quality.
–
CCB7
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing "E" accepts the new calibration data.
Reject
: The data is not accepted, a recalibration won’t
be started.
Recalibrate
: The data is rejected and a new calibration
will be started.
E1
CCB8
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that
measurement can take place?
–
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CC
Calibration with
rel. data entry
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
–
CCC1
1. (0...30%):
1. Voltage
2. (70...100%)
2. Voltage
20 %
0600mV
80 %
-0600mV
Enter calibration points
In this field, create two measured value pairs (pair 1
and pair 2).
Measured value pair 1 in the range 0...30 %: assign, for
example, the voltage 0600 mV to the percentage value
20 %.
Measured value pair 2 in the range 70...100 %: assign,
for example, the voltage –0600 mV to the percentage
value 80 %.
The settings made become effective immediately after
confirmation with "E".
E2
CCC3
Offset too high / too low
Note in display:
Error message if the entered offset leaves the
maximum range.
–
CCC4
Accept
Reject
Recalibrate
End of calibration
Accept: Pressing "E" accepts the new calibration data.
Reject: The data is not accepted, a recalibration won’t
be started.
Recalibrate: The data is rejected and a new calibration
will be started.
E1
CCC5
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that
measurement can take place?
–