TOC Analyzer CA79
Commissioning
Hauser
33
1. Press the
Interval list
key to open the interval editor.
The interval sequence can be edited in the editor once you press the
Interval list
key.
A0046974
24
Editor
The time is the measuring time with the rinse duration in minutes.
If
NA
is selected for a port in a step or if the list is completely full, the sequence starts
again with step 1 in the measuring mode so that the samples are monitored
continuously.
The system creates a new data file after every step. This differs from the continuous mode
without an interval where a new file is created after 24 hours at the very latest if the
measurement process has not been interrupted.
The first row must contain a sample and a time.
Software adapted to CFR 21 Part 11
CFR 21 Part 11 is part of Title 21 of the Code of Federal Regulations, which sets down the
rules of the United States Food and Drug Administration (FDA) regarding electronic
records and electronic signatures (ERES). Part 11, as it is commonly known, defines the
criteria under which electronic records and electronic signatures are considered to be
trustworthy, reliable and equivalent to paper records (Title 21 CFR Part 11 Section 11.1
(a)).
A new user management function was introduced to customize the analyzer software for
Part 11. In the software program that is based on CRF 21 Part 11, the
User management
menu is displayed in the main menu.