Calibration curve
1
Factors and R
2
2
Calibration curve
3
List of measured values
It is recommended to repeat the measurement series three times. The standard
volumes used are 500 ml and suffice for several measurements. The measurement
results should be constant within this measurement series and be close to one
another.
System suitability test (SST)
The measurement must be stopped for the system suitability test (SST).
A0046957
16
"Provide SST solutions" message
The system must be switched on one hour before the SST starts so that a suitable
operating temperature is reached. The SST solutions must be warmed to at least room
temperature beforehand. If temperatures below 18 °C are displayed at the start of the
measurement, the measurement must be stopped until the SST solutions have at least
reached room temperature. The optimum temperature range is between 20 °C and 25
°C as the starting temperature for the SST measurement.