TTR31, TTR35
Trouble-shooting
Hauser
29
8.2 Repair
A repair is not planned.
8.3 Disposal
Please pay particular attention to the local disposal regulations of your country. When disposing,
ensure that the materials of the device components are separated and processed accordingly.
E044
Output current drifts too much
(± 0.5 mA)
Return device to E+H
Code
Explanation
Remedy
W107
Simulation active
Switch off the output simulation for
output 1 and output 2
W202
Measured value outside of the sensor
range
Operate the device in the specified
temperature range
W209
Device starts
W210
Configuration modified (warning code
will be displayed for 15 s approx.)
W212
Sensor signal outside the permitted range
Operate the device in the specified
temperature range
W250
Number of switch cycles exceeded
Replace the device
W270
Short-circuit or overload at output 1
Check output wiring.
Extend the load resistance at output 1
W280
Short-circuit or overload at output 2
Check output wiring.
Extend the load resistance at output 2
Code
Explanation
Remedy