Technical data
StamoLys CA 71 SI
42
Hauser
10
Technical data
10.1
Input
10.2
Output
10.3
Power supply
Measured variable
SiO
2
[µg/l / mg/l]
Measuring range
1 ... 200 µg/l (SI-A)
0.01 ... 5.00 mg/l (SI-B)
Wave length
810 nm
Reference wave length
565 nm (SI-B only)
Output signal
0/4 ... 20 mA
Signal on alarm
Contacts: 2 limit contacts (per channel), 1 system alarm contact
optionally:end of measurement (with two channel version display of
channel no. available)
Load
max. 500
Ω
Serial interface
RS 232 C
Load capacity
230 V / 115 V AC max. 2 A, 30 V DC max. 1 A
Supply voltage
115 V AC / 230 V AC ±10%, 50/60 Hz
Power consumption
approx. 50 VA
Current consumption
ca. 0.2 A at 230 V
ca. 0.5 A at 115 V
Fuses
1 x time-lag 0.5 A for electronics
2 x medium time-lag 0.2 A for photometer
1 x time-lag 0.1 A for motors
Summary of Contents for Stamolys CA71CL