Maintenance and Troubleshooting
Hardware Installation and Maintenance Manual
B
–
21
SCS status check
These steps should be performed only if there is a suspected problem with the
analyzer.
1.
Open the SCS door.
2.
Read and record the flow meter settings while the gas is flowing.
3.
Close the SCS door.
4.
Compare the current readings with the past readings to determine
any variations. Reading levels should remain consistent.
5.
If reading levels decrease, check the filters.
To check the filter
1.
Shut down the system following the procedure in
2.
Inspect, repair or replace the filter as required. Refer to
3.
Restart the system.
a.
Open the low pressure flare or atmospheric vent header shut-
off valve for the effluent from the sample bypass.
b.
Open the sample supply shut-off valve to the SCS.
Performing this check will cause reading errors for up to 4 hours
while the enclosure temperature stabilizes.
Do not leave the SCS door open any longer than absolutely
necessary. SpectraSensors recommends no more than 60
seconds.
It is not necessary to turn off system power to perform this
function.
Contact Service for additional information, refer to
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...