If DHCP is selected, the
IP
and
Subnet Mask
will be assigned by the
DHCP server and therefore are not editable.
•
If Static is selected, the
IP
and
Subnet Mask
must be entered
manually. Contact your local Network Administrator for assistance if
input data needs to be confirmed.
•
Port
specifies the network port under which the Modbus application
takes place. The default value for most Modbus applications is 502.
•
The
ID
, which is used with the Modbus communication, can be set to
any value between 1 and 32.
All settings are applied by pressing
.
Analog Output Settings (Analogues) Menu Options
From the MAIN MENU, press Analogues to access the 4-20mA INTERFACE
SETTINGS, 4-20mA VALUES, CONCENTRATION ALARM RELAY (LS2) and
4-20mA CALIBRATION screens.
4-20mA interface settings
The 4-20mA INTERFACE SETTINGS can be accessed through the ANALOGUES
menu. When accessed, the following screen displays (refer to Figure 4–53).
Figure 4–52
Digitals - TCP/IP settings