Electrical connection
Smart System for Aquacultures SSP200B
18
Hauser
6.3
Cable entries and terminals
H
5
6
7
8
I
G
F
E
D
C
A
1
2
3
4
B
A0018025
12
Underside of housing with labeled cable entries and terminals
Labeling on underside of housing Description for Liquiline CM444
Description for Modbus Device
Edge
1 to 8
Sensors 1 to 8
Do not use
A
Supply voltage
Do not use
B
RS485 In or M12 DP/RS485 (not
used in factory configuration)
Do not use
C
Freely usable
LTE antenna
D
Current outputs and inputs, relays
(not used in factory configuration)
LTE antenna
E
Do not use
Do not use
F
Current outputs and inputs, relays
(not used in factory configuration)
Ethernet
G
Current outputs and inputs, relays
(not used in factory configuration)
Supply voltage
H
Freely usable
Do not use
I
RS485 Out and M12 Ethernet
(M12 Ethernet used in factory
configuration)
Do not use
6.4
Electrical connection Liquiline CM444
NOTICE
The transmitter does not have a power switch!
‣
Provide a protected circuit breaker in the vicinity of the transmitter at the place of
installation.
‣
The circuit breaker must be a switch or power switch, and must be labeled as the circuit
breaker for the device.