Prosonic Flow 92
3 Installation
Hauser
23
3.3.3
Installing the measuring sensors Prosonic Flow U
Intermediate step:
Preparation of tensioning band lock for initial mounting.
1.
Guide the tensioning band with the Velcro side face up through the lowest opening
in the tensioning band fastener (a).
2.
Loop the tensioning band back through the top opening (b).
3.
Twist the tensioning band in such a way that you can guide it through the centre
opening (C) to such an extent that you have at least 20 mm for pressing the two
Velcro sides together (d).
4.
Pull the tensioning band back tightly through the bottom opening (e).
5.
Continue with the sensor installation.
Fig. 18:
Preparing the tensioning band lock
1.
Loop the tensioning band for U sensors around the pipe and feed it through the
lock. Make sure the band is not twisted and that the smooth surface is oriented
towards the pipe. Do not yet tighten the band. Continue with the second band.
Fig. 19:
Preparing the tensioning bands for sensor installation
2.
Prepare the U sensor for installation:
Use the sensor installation Site Setup menu (see Page 67) to obtain the appropriate
sensor distance for your application.
!
Note!
The U sensor only supports the 2 traverses configuration. Make sure that the number of
traverses is set to 2 (see Page 71) in the “Site Data” menu.
F06-92
x
x
x
x
x
x
-17-x
x
-06-x
x
-001
F
0
6
-9
2
x
xxxxx
-1
7
-0
5
-0
6
-xx-0
0
2
Summary of Contents for prosonic flow 92
Page 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Page 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Page 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Page 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Page 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Page 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...