Prosonic Flow 92
3 Installation
Hauser
15
3.2.3
Orientation
Vertical orientation
Recommended orientation with upward direction of flow (View A). Entrained solids sink
down. Gases rise away from the measuring sensor when fluid is not flowing. The piping
can be completely drained and protected against build-up.
Horizontal orientation
In the recommended installation range (C) with horizontal orientation (View B), gas and
air collections at the top of the pipe and problematic deposits at the bottom of the pipe
have a minor influence on the measurement.
Fig. 6:
Orientation (A = Vertical, B = Horizontal, C = Recommended installation range max. 120°)
3.2.4
Inlet and outlet runs
If possible, install the sensor well clear of assemblies such as valves, T-pieces, elbows,
etc. If several flow obstructions are installed, the longest inlet or outlet run must be con-
sidered. Compliance with the following requirements for the inlet and outlet runs is rec-
ommended to ensure measuring accuracy:
Fig. 7:
Inlet and outlet runs
1 = Valve; 2 = Pump; 3 = Two pipe bends in different directions
3.2.5
Connecting cable length
Shielded cables are available in the following lengths:
5 m and 10 m
"
Caution!
Route the cable well clear of electrical machines and switching elements.
F06-9x
x
x
x
x
x
x
-11
-05
-0
0-x
x
-004
≥
15 x DN
≥
40 x DN
≥
20 x DN
≥
40 x DN
≥
5 x DN
1
2
3
F06-9xxx
xxxx-
1
1
-05-00-xx
-003
Summary of Contents for prosonic flow 92
Page 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Page 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Page 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Page 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Page 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Page 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...