Proline Prosonic Flow 90
9 Trouble-shooting
Hauser
91
9.8
Installation/removal of flowmeter sensors W
“Insertion”
The active part of the flowmeter sensor W “Insertion” can be replaced without interrupting the
process.
1.
Pull the sensor connector (1) out of the sensor cover (3).
2.
Remove the small retainer ring (2). This is located on the top of the sensor neck and keeps the
sensor cover in place.
3.
Remove the sensor cover (3) and spring (4).
4.
Remove the large retainer ring (5). This keeps the sensor neck (6) in place.
5.
The sensor neck can now be pulled out. Note that you must reckon with a certain amount of
resistance.
6.
Pull the sensor element (7) out of the sensor holder (8) and replace it with a new one.
7.
Installation is the reverse of the removal procedure.
A0001147
Fig. 51:
Installation/removal of flowmeter sensor W “Insertion”
1
Sensor connector
2
Small retainer ring
3
Sensor cover
4
Spring
5
Large retainer ring
6
Sensor neck
7
Sensor element
8
Sensor holder
1
2
3
4
5
6
7
8
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...