Commissioning
Proservo NMS83
98
Hauser
9.5.4
Configuration of a connected RTD
1. Make sure the RTD is connected as defined by the terminal assignment → 46.
2. Navigate to the submenu of the respective Analog I/O module: Setup → Advanced
setup → Input/output → Analog IP
3. Go to the
RTD type
parameter (→ 191) and specify the type of the connected
RTD.
4. Go to the
RTD connection type
parameter (→ 192) and specify the type of
connection of the RTD (2-, 3- or 4-wire).
5. Go to the
Input value
parameter and check whether the indicated temperature
matches the actual temperature.
6. Go to the
Minimum probe temperature
parameter (→ 193) and specify the
minimum approved temperature of the connected RTD.
7. Go to the
Maximum probe temperature
parameter (→ 194) and specify the
maximum approved temperature of the connected RTD.
8.
2
1
3
A0029269
1
Datum plate
2
RTD
3
Probe position (→ 194)
Go to the
Probe position
parameter and enter the mounting position of the RTD
(measured from the datum plate).
This parameter, in conjunction with the measured level, determines whether the
measured temperature refers to the product or to the gas phase.