Commissioning
Proservo NMS83
92
Hauser
Interface profile measurement
Setting interface profile procedure
The interface profile operation measures a profile starting at the upper interface level up
to the liquid level.
1. Navigate to: Setup → Advanced setup → Sensor config → Profile density → Profile
density offset distance
2. Input the desired value for the
Profile density offset distance
parameter.
The value of the profile density offset distance defines the distance between the
start point (tank bottom in this case) and the first measurement point.
3. Navigate to: Setup → Advanced setup → Sensor config → Profile density → Profile
density interval
4. Input the desired value for the
Profile density interval
parameter.
This completes setting interface profile procedure.
2
3
4
A
C
B
1
5
6
A0029110
47
Interface profile movement (The numbers show the order of the displacer movement.)
A
Profile density interval
B
Profile density offset distance
C
Tank profile range