Proservo NMS83
Commissioning
Hauser
91
Tank profile measurement
Setting tank profile procedure
The tank profile operation measures a profile starting at the physical tank bottom up to
the liquid level.
1. Navigate to: Setup → Advanced setup → Sensor config → Profile density → Profile
density offset distance
2. Input the desired value for the
Profile density offset distance
parameter.
The value of the profile density offset distance defines the distance between the
start point (tank bottom in this case) and the first measurement point.
3. Navigate to: Setup → Advanced setup → Sensor config → Profile density → Profile
density interval
4. Input the desired value for the
Profile density interval
parameter.
This completes the setting tank profile procedure.
2
3
5
6
7
A
B
C
D
1
4
A0029108
46
Tank profile movement (The numbers show the order of the displacer movement.)
A
Profile density interval
B
Profile density offset distance
C
Datum plate
D
Tank profile range