Proservo NMS83
Operability
Hauser
63
7.3.9
Access code and user roles
Meaning of the access code
An access code can be defined in order to distinguish between the following user roles:
User role
Definition
Maintenance
• Knows the access code.
• Has write access to all parameters (except service parameters).
Operator
• Doesn't know the access code.
• Has write access to only a few parameters.
• The description of parameters states which role is needed at least for read and write
access to each parameter → 153.
• The current user role is indicated by the
Access status display
parameter.
• If the access code is
"0000"
, every user is in the
Maintenance
role. This is the
default setting on delivery of the device.
Defining an access code
1. Navigate to: Setup → Advanced setup → Administration → Define access code
→ Define access code
2. Enter the intended access code (max. 4 digits).
3. Repeat the same code in the
Confirm access code
parameter.
The user is in the
Operator
role. The -symbol appears in front of all write-
protected parameters.
Switching to the "Maintenance" role
If the -symbol appears on the local display in front of a parameter, the parameter is
write-protected because the user is in the
Operator
role. To switch to the
Maintenance
role, proceed as follows:
1. Press .
The input prompt for the access code appears.
2. Enter the access code.
The user is in the
Maintenance
role. The -symbol in front of the parameters
disappears; all previously write-protected parameters are now re-enabled.
Switching back to the "Operator" role automatically
The user automatically switches back to the
Operator
role:
• if no key is pressed for 10 minutes in the navigation and editing mode.
• 60 s after going back from the navigation and editing mode to the standard view
(measured value display).
Switching back to the "Operator" role manually
1. Navigate to: Setup → Advanced setup → Enter access code
2. Enter any number other than the previously defined access code.
The user is in the
Operator
role. The -symbol appears in front of all write-
protected parameters.