Operating menu
Proservo NMS83
176
Hauser
"Sensor calibration" wizard
Navigation
Setup → Calibration → Sensor cal.
Sensor calibration
Navigation
Setup → Calibration → Sensor cal. → Sensor cal.
Description
This sequence calibrates the sensor of the servo.
Additional information
Read access
Operator
Write access
Maintenance
Offset weight
Navigation
Setup → Calibration → Sensor cal. → Offset weight
Description
Sets the weight that is used for the lower point sensor calibration. Changing the value will
delete the calibration data.
User entry
0 to 150 g
Factory setting
Dependent on the device version
Additional information
Read access
Operator
Write access
Maintenance
For density measurement application, it is recommended to apply 50 g.
Span weight
Navigation
Setup → Calibration → Sensor cal. → Span weight
Description
Sets the weight that is used for the middle point sensor calibration. Changing the value
will delete the calibration data.
User entry
10 to 999.9 g
Factory setting
Dependent on the device version
Additional information
Read access
Operator
Write access
Maintenance