Commissioning

Proline t-mass T 150

42

Hauser

Device tag

Enter the name for the measuring point.

Temperature

Displays the temperature currently measured.

Pipe inner diameter

Enter the inner diameter of the pipe.

Installation factor

Enter the factor to adjust the installation conditions.

Assign status input

Select the function for the status input.

Assign current output

Select process variable for current output.

4 mA value

Enter 4 mA value.

20 mA value

Enter 20 mA value.

Operating mode

Specify the output as a pulse, frequency or switch output.

Assign frequency output

Select the process variable for the frequency output.

Measuring value at minimum frequency

Enter the measured value at the minimum frequency.

Measuring value at maximum frequency

Specify the measured value at maximum frequency.

Switch output function

Select the function for the switch output.

Assign limit

Select the process variable for the limit function.

Switch-off value

Enter the measured value for the switch-off value.

Switch-on value

Enter the measured value for the switch-on value.

Assign status

Select the device status for the switch output.

Assign diagnostic behavior

Select the diagnostic behavior for the switch output.

Assign pulse output

Select the process variable for the pulse output.

Value per pulse

Enter the measured value for the pulse output.

10.5 

Defining the tag name

To enable fast identification of the measuring point within the system, you can enter a unique

designation using the 

Device tag

 parameter and thus change the factory setting.

Navigation

"Setup" menu → Device tag

Parameter overview with brief description

Parameter

Description

User entry

Factory setting

Device tag

Enter tag for measuring
point.

Max. 32 characters, such as
letters, numbers or special
characters (e.g. @, %, /).

t-mass

10.6 

Protecting settings from unauthorized access

The following options exist for protecting the configuration of the measuring device from

unintentional modification after commissioning:

• Write protection via access code

• Write protection via write protection switch

• Write protection via keypad lock

10.6.1 

Write protection via access code

With the customer-specific access code, the parameters for the measuring device

configuration are write-protected and their values can no longer be changed via local

operation.

Summary of Contents for Proline t-mass T 150

Page 1: ...se Instructions are Brief Operating Instructions they do not replace the Operating Instructions included in the scope of supply For detailed information refer to the Operating Instructions and other d...

Page 2: ...1 Installation conditions 13 6 2 Mounting the measuring device 21 6 3 Post installation check 24 7 Electrical connection 25 7 1 Connection conditions 25 7 2 Connecting the measuring device 28 7 3 Ensu...

Page 3: ...rical symbols Symbol Meaning A0011197 Direct current A terminal to which DC voltage is applied or through which direct current flows A0011198 Alternating current A terminal to which alternating voltag...

Page 4: ...ocedures processes or actions that are preferred A0011184 Forbidden Indicates procedures processes or actions that are forbidden A0011193 Tip Indicates additional information A0011194 Reference to doc...

Page 5: ...e Application and media The measuring device described in these Operating Instructions is intended only for flow measurement of liquids Measuring devices for use in hazardous areas in hygienic applica...

Page 6: ...rifying the corrosion resistance of fluid wetted materials but does not accept any warranty or liability as minute changes in the temperature concentration or level of contamination in the process can...

Page 7: ...ific EC Declaration of Conformity Endress Hauser confirms this by affixing the CE mark to the device 2 6 IT security We only provide a warranty if the device is installed and used as described in the...

Page 8: ...ss Hauser 3 Product description 3 1 Product design 1 2 2 2 2 3 2 4 2 5 2 6 2 7 7 Es c 1 2 3 4 5 6 8 A0017196 1 Electronics compartment cover 2 Display module 3 Terminal block 4 Electronics module 5 Ca...

Page 9: ...A0015502 A0013699 Do the nameplate data match the ordering information on the delivery note A0015502 A0013697 Is the CD ROM with the Technical Documentation depends on device version and documents pre...

Page 10: ...r no 1 2 3 4 A0021952 1 Example of a nameplate 1 Order code 2 Serial number Ser no 3 Extended order code Ext ord cd 4 2 D matrix code QR code For detailed information on the breakdown of the specifica...

Page 11: ...al number Ser no of the device see nameplate 1 10 All the associated documentation is displayed Via the 2 D matrix code QR code 1 Launch the Endress Hauser Operations App 2 Scan the 2 D matrix code QR...

Page 12: ...ransport Proline t mass T 150 12 Endress Hauser 5 3 Packaging disposal For detailed information about disposing of the packaging materials refer to the Operating Instructions for the device on the CD...

Page 13: ...nformation on aligning with the flow direction 19 Installation is generally not recommended in the event of high vibrations or unstable internal fittings Orientation Recommendation Vertical orientatio...

Page 14: ...nstallation As a result the insertion depth is fixed after initial installation and the clamping ferrules can no longer be replaced Pay attention to information on preconditions and on determining the...

Page 15: ...the permitted height B C may not be greater than 53 mm 2 09 in Determining the insertion depth before mounting for the first time For all nominal diameters 8 B D 1 Controlling the insertion depth afte...

Page 16: ...l coupling process connection with a self centering sealing ring is used determine seal thickness E 4 If an aseptic liner or a conical coupling process connection is used set the seal thickness E to e...

Page 17: ...mum of 2 decimal places Example 17 05 in When ordering a maximum of 3 decimal places can be declared NOTICE Certain dimensions are required for determining the customized length Pay attention to infor...

Page 18: ...E may not be greater than 77 mm 3 03 in 6 If an aseptic liner or a conical coupling process connection is used set the seal thickness E to equal zero and do not take it into consideration NOTE The pip...

Page 19: ...e is aligned at a 90 angle to the direction of flow as shown in the graphic 2 Rotate the sensor so that the arrow marking on the sensor body corresponds to the direction of flow 3 Align the scale to t...

Page 20: ...0 C 4 to 140 F the readability of the display may be impaired at temperatures outside the temperature range If operating outdoors Avoid direct sunlight particularly in warm climatic regions System pre...

Page 21: ...ion layer is Order code for Insertion Length option L5 110mm 4 100 mm 3 94 in The following is recommended for thicker insulation layers Order code for Insertion Length option L6 330mm 13 320 mm 12 6...

Page 22: ...process connection Tighten thread adapter nut with x turns For PEEK clamping ferrules continue with Step 4 For metallic clamping ferrules continue with Step 5 For hygienic process connections continu...

Page 23: ...cable entries do not point upwards A0013964 6 2 4 Turning the transmitter housing To provide easier access to the connection compartment or display module the transmitter housing can be turned clockw...

Page 24: ...sensor been selected 13 According to sensor type According to medium properties According to medium temperature According to process pressure Does the arrow on the sensor match the direction of flow o...

Page 25: ...irements Electrical safety In accordance with applicable federal national regulations Permitted temperature range 40 C 40 F to 80 C 176 F Minimum requirement cable temperature range ambient temperatur...

Page 26: ...se frequency switch output 4 Signal transmission 4 20 mA HART 5 Ground terminal for cable shield Supply voltage Order code for Power supply Terminal numbers 1 L 1 2 L 1 Option D DC 18 to 30 V 1 Secure...

Page 27: ...communication types on the device side 1 2 4 3 5 A0016809 Pin Assignment Coding Plug socket 1 L DC24 V A Plug 2 Status input 3 Status input 4 L DC24 V 5 Grounding shielding 4 20 mA HART with pulse fr...

Page 28: ...d be defeated Use suitable cable glands corresponding to the degree of protection If measuring device is delivered without cable glands Provide suitable cable gland for corresponding connecting cable...

Page 29: ...icant The threads on the cover are coated with a dry lubricant Reverse the removal procedure to reassemble the transmitter For HART communication When connecting the cable shielding to the ground term...

Page 30: ...inspection Are the power supply and signal cables correctly connected Does the supply voltage correspond to the specifications in the connection diagram Do the cables comply with the requirements Do t...

Page 31: ...rs and maintenances Operator Maintenance task oriented function oriente d Expert A0014058 EN 6 Schematic structure of the operating menu 8 1 2 Operating philosophy The individual parts of the operatin...

Page 32: ...ring channel number symbol for event behavior 1 4 Status area 1 5 Measured value 1 6 Unit for the measured value 1 7 Operating elements 2 Operational display with measured value shown as 1 bar graph 1...

Page 33: ...ed variables Symbol Meaning A0013711 Volume flow A0013710 Mass flow A0013947 Temperature A0013943 Totalizer A0013945 Current output A0017270 Status input Symbols for measurement channel numbers A00163...

Page 34: ...s A0013967 A0013968 A0013972 A0013963 Submenu Wizard Parameters within a wizard Parameter locked 8 2 3 Editing view Input mask Operating symbols in the numeric editor A0013985 A0016621 A0013986 Confir...

Page 35: ...the selection bar downwards in a choose list With a Wizard Confirms the parameter value and goes to the next parameter With a text and numeric editor Moves the selection bar to the right forwards in...

Page 36: ...ress the keys simultaneously Reduces the contrast brighter setting A0013954 Plus Enter key combination press and hold down the keys simultaneously Increases the contrast darker setting A0013955 Minus...

Page 37: ...which the user is currently logged on is indicated by the Access status display parameter Navigation path Operation menu Access status display parameter 8 2 7 Disabling write protection via access cod...

Page 38: ...message Keylock off appears on the display The keypad lock is switched off 8 3 Access to the operating menu via the operating tool For detailed information about access to the operating menu via oper...

Page 39: ...tions for remote operation via HART protocol 1 Control system e g PLC 2 Field Communicator 475 3 Computer with operating tool e g FieldCare AMS Device Manager SIMATIC PDM 4 Commubox FXA195 USB 5 Field...

Page 40: ...ee the Operating Instructions for the device 10 10 Commissioning 10 1 Function check Before commissioning the device make sure that the post installation and post connection checks have been performed...

Page 41: ...e Factory setting English or ordered local language XXXX XXX X X 20 50 Operation Setup Main menu 0104 1 Language English Espa ol Fran ais Language English Deutsch 0104 1 Espa ol Fran ais Language Engl...

Page 42: ...switch on value Assign status Select the device status for the switch output Assign diagnostic behavior Select the diagnostic behavior for the switch output Assign pulse output Select the process vari...

Page 43: ...w The device locks the write protected parameters automatically after 60 s if the user skips back to the operational display mode from the navigation and editing view If write access is activated via...

Page 44: ...parameter the Hardware locked option is displayed In addition on the local display the symbol appears in front of the parameters in the header of the operational display and in the navigation view 4...

Page 45: ...current output settings A0022311 EN 8 Message for remedial measures 1 Diagnostic information 2 Short text 3 Service ID 4 Diagnostic behavior with diagnostic code 5 Operation time of occurrence 6 Reme...

Page 46: ......

Page 47: ......

Page 48: ...www addresses endress com...

Reviews: