Proline t-mass A 150 HART
Commissioning
Hauser
67
10.5.8 Configuring the totalizer
You can configure the totalizer in the
Totalizer
submenu.
Navigation path
"Setup" menu → Advanced setup → Totalizer
Structure of the submenu
Totalizer
→
Assign process variable
Unit
Failure mode
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection/
User entry
Factory setting
Assign process
variable
-
Select process
variable for
totalizer.
Result
The selection
determines the
choose list of the
Unit
parameter.
• Off
• Mass flow
• Corrected
volume flow
• FAD volume
flow
Mass flow
Unit
One of the following options
is selected in the
Assign
process variable
parameter:
• Mass flow
• Corrected volume flow
• FAD volume flow
Select the unit for
the process
variable of the
totalizer.
Unit choose list
Country-
dependent:
• kg
• lb
Failure mode
One of the following options
is selected in the
Assign
process variable
parameter:
• Mass flow
• Corrected volume flow
• FAD volume flow
Define how the
totalizer behaves
in an alarm
condition.
• Stop
• Actual value
• Last valid value
Stop
10.5.9 Configuring the local display
Navigation path
"Setup" menu → "Advanced setup" menu →"Display" menu
Parameter overview with brief description
Parameter
Description
Selection/
User entry
Factory setting
Format display
Select how measured
values are shown on the
display.
• 1 value, max. size
• 1 ba 1 value
• 2 values
• 1 value large + 2 values
• 4 values
1 value, max. size