Proline t-mass A 150 HART
Installation
Hauser
17
Recommended inlet and outlet runs (without flow conditioner)
15 × DN
2 × DN
1
4
2
6
3
15 × DN
2 × DN
15 × DN
2 × DN
20 × DN
2 × DN
35 × DN
2 × DN
5
50 × DN
2 × DN
A0016942
1
reduction
2
expansion
3
90° elbow or T-piece
4
2 × 90° elbow
5
Control valve
6
2 × 90° elbow (3-dimensional)
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
Outlet run for pressure or temperature transmitter
If a pressure or temperature measuring device is installed downstream of the measuring
device, make sure there is sufficient distance between the two devices.
PT
2…5 × DN
TT
A0015548
PT Pressure measuring device
TT Temperature measuring device
Flow conditioner (19 hole) for use with fixed flanges
If the inlet runs cannot be observed, the use of a flow conditioner is recommended.
2 × DN
5 × DN
A0015547
3
Recommended inlet and outlet runs when using a flow conditioner
Summary of Contents for Proline t-mass A 150