Proline t-mass 65 PROFIBUS DP/PA
Commissioning
Hauser
55
7.3.4
"Pressure" Quick Setup menu
The individual process pressure for every gas group can be programmed with this Quick
Setup. If only one gas group is being used, then only the function PROCESS PRESSURE 1
needs to be programmed, PROCESS PRESSURE 2 can remain with default settings.
A0009908-en
!
Note!
• The measuring device operates with absolute pressure only. All gauge pressure must be
converted to absolute pressure.
• If a pressure compensating input is being used, then the input signal value overrides the
manually programmed value. The pressure input value applies to both gas groups. i.e. 2
independent pressure values are no longer possible.
• The Quick Setup Gas function is not available if an in-situ calibration function has been
performed on the measuring device as the in-situ calibration curve refers to the sensor
power at each recorded flow point. Therefore, the programmed pressure settings become
redundant
+
+