Proline Prosonic Flow P
Installation
Hauser
27
Material
The following material is required for mounting:
• Two strapping bands incl. mounting bolts and centering plates where necessary (already
• Two measuring wires, each with a cable lug and a fixer to fix the strapping bands
• Two sensor holders
• Coupling medium (coupling pad or coupling gel) for an acoustic connection between the
sensor and pipe
• Two sensors incl. connecting cables
Installation is unproblematic up to DN 400 (16"), as of DN 400 (16") check the distance
and angle (180°) diagonally with the wire length.
Procedure:
1. Prepare the two measuring wires: arrange the cable lugs and fixer such that the distance
they are apart corresponds to the wire length (SL). Screw the fixer onto the measuring
wire.
1
2
3
4
5
6
7
8
9
10
11
SL
A0043379
15
Fixer and cable lugs at a distance that corresponds to the wire length (SL)
2. With measuring wire 1: fit the fixer over the mounting bolt of strapping band 1 that is
already securely mounted. Run measuring wire 1 clockwise around the measuring pipe.
Fit the cable lug over the mounting bolt of strapping band 2 that can still be moved.
3. With measuring wire 2: fit the cable lug over the mounting bolt of strapping band 1 that
is already securely mounted. Run measuring wire 2 counterclockwise around the
measuring pipe. Fit the fixer over the mounting bolt of strapping band 2 that can still be
moved.