Installation
Proline Prosonic Flow P
22
Hauser
12. Only slightly tighten strapping band 2 for final mounting so that it can still be adjusted.
The distance/offset from the center of strapping band 2 to the center of strapping band
1 is indicated by the sensor distance of the device.
13. Align strapping band 2 so that it is perpendicular to the measuring pipe axis and
parallel to strapping band 1.
14. Position the two strap bolts on tensioning strap 2 on the measuring pipe so they are
parallel to one another and offset at the same height/clock position (e.g. 10 and 4
o'clock) in relation to the two strap bolts on tensioning strap 1. A line drawn on the
measuring pipe wall that is parallel to the measuring pipe axis can be helpful here. Now
set the distance between the center of the strap bolts at the same level so that it exactly
matches the sensor distance. An alternative method is to use the wire length → 27.
15. Tighten strapping band 2 so that it cannot slip.
L
WARNING
Risk of injury!
‣
To avoid sharp edges, trim the cut edges after shortening the strapping bands. Wear
suitable gloves and protective goggles.
1
2
3
≥ 500 mm (20 in)
3
A0043374
10
Holder with strapping bands (large nominal diameters)
1
Strap bolt with guide*
2
Strapping band*
3
Tensioning screw
Summary of Contents for Proline Prosonic Flow P
Page 39: ......