Proline Promass Q 500 PROFIBUS PA
Installation
Hauser
23
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
tube, avoid the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical
pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the
nominal diameter prevent the sensor running empty while measurement is in progress.
1
2
3
4
5
A0028773
6
Installation in a down pipe (e.g. for batching applications)
1
Supply tank
2
Sensor
3
Orifice plate, pipe restriction
4
Valve
5
Batching tank
DN
Ø orifice plate, pipe restriction
[mm]
[in]
[mm]
[in]
25
1
14
0.55
50
2
28
1.10
80
3
50
1.97
100
4
65
2.60
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction (direction of medium flow through the piping).
Orientation
Recommendation
A
Vertical orientation
A0015591
B
Horizontal orientation, transmitter at
top
A0015589
Exceptions:
Summary of Contents for Proline Promass Q 500
Page 279: ......