Proline Promass F 200 HART
Mounting
Hauser
21
Vibrations
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by plant vibrations.
6.1.3
Special mounting instructions
Outlet run for peripheral device
If a pressure and temperature measuring device are installed downstream from the measuring device,
make sure there is sufficient distance between the two devices.
3…5 × DN
PT
TT
4…8 × DN
A0016893
PT
Pressure measuring device
TT
Temperature measuring device
Rupture disk
Make sure that the function and operation of the rupture disk is not impeded through the installation
of the device. The position of the rupture disk is indicated on a sticker beside it.
The existing connecting nozzles are not designed for a rinse or pressure monitoring function.
L
H
45°
RUPTURE DISK
i
2
3
1
F
G
E
A0008361
1
Indication label for the rupture disk
2
Rupture disk with 1/2" NPT internal thread with 1" width across flat
3
Transport protection
DN
E
F
G
H
L
[mm]
[in]
[mm]
[in]
[in]
[in]
[mm]
[in]
[mm]
[in]
8
3/8
Approx.42
Approx.1.65
AF 1
½ NPT
62
2.44
216
8.50
15
1/2
Approx.42
Approx.1.65
AF 1
½ NPT
62
2.44
220
8.66
25
1
Approx.42
Approx.1.65
AF 1
½ NPT
62
2.44
260
10.24