Technical data
Proline Promass F 100 Modbus RS485
96
DN
x
[mm]
[in]
[kg/m
3
]
40
1½
90
50
2
90
80
3
110
100
4
130
150
6
200
250
10
200
Calculation example for gas
• Sensor: Promass F, DN 50
• Gas: Air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)
• Measuring range (liquid): 70
000 kg/h
• x = 90 kg/m³ (for Promass F, DN 50)
Maximum possible full scale value:
max(G)
=
max(F)
· ρ
G
: x = 70
000 kg/h · 60.3 kg/m³ : 90 kg/m³ = 46
900 kg/h
Recommended measuring range
"Flow limit" section → 108
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value do not override the electronics unit, with the
result that the totalizer values are registered correctly.
16.4 Output
Output signal
Modbus RS485
Physical interface
In accordance with EIA/TIA-485-A standard
Terminating resistor
• For device version used in non-hazardous areas or Zone 2/Div. 2: integrated
and can be activated via DIP switches on the transmitter electronics module
• For device version used in intrinsically safe areas: integrated and can be
activated via DIP switches on the Safety Barrier Promass 100
Signal on alarm
Depending on the interface, failure information is displayed as follows: