Proline Promass E 100 Modbus RS485
Commissioning
Hauser
61
10.5.1 Parameter overview with brief description
Parameter
Prerequisite
Description
Selection/
User entry
Factory setting
Assign simulation
process variable
-
Select a process
variable for the
simulation process
that is activated.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
The range of
options
increases if
the measuring
device has one
or more
application
packages.
Off
Value process
variable
A process variable is
selected in the
Assign simulation
process variable
parameter.
Enter the simulation
value for the
selected process
variable.
Depends on the
process variable
selected
-
Simulation device
alarm
-
Switch the device
alarm on and off.
• Off
• On
Off
10.6 Protecting settings from unauthorized access
The following option exists for protecting the configuration of the measuring device from
unintentional modification after commissioning: Write protection via write protection
switch
10.6.1 Write protection via write protection switch
The write protection switch makes it possible to block write access to the entire operating
menu with the exception of the following parameters:
• External pressure
• External temperature
• Reference density
• All parameters for configuring the totalizer
The parameter values are now read only and cannot be edited any more:
• Via service interface (CDI)
• Via Modbus RS485