Proline Promag W 400
Operation options
Hauser
51
1
2
3
4
1
2
3
4
1
2
ON OFF Default Ethernet network settings
IP 192.168.1.212
Write protection
128
64
32
16
8
4
2
1
IP
Address
setting
(last octet)
A0023353
• Once the DIP switch has been activated, the device must be restarted
before the device uses the standard IP address.
• If the standard IP address (top DIP switch No. 2 = ON) is used, there is
no connection to the EtherNet/IP network.
Software of the computer
Web browsers supported
• Microsoft Internet Explorer (min. 8.x)
• Mozilla Firefox
• Google chrome
Recommended operating systems
• Windows XP
• Windows 7
User rights for TCP/IP settings
User rights required for TCP/IP settings (e.g. for changes to IP address, subnet
mask)
Computer configuration
• JavaScript is enabled
• If JavaScript cannot be enabled, enter http://192.168.1.212/basic.html in the
address line of the Web browser. A fully functional but simplified version of
the operating menu structure starts in the Web browser.
8.3.3
Establishing a connection
Configuring the Internet protocol of the computer
IP address
192.168.1.XXX; for XXX all numerical values except: 0, 212 and 255 → e.g.
192.168.1.213
Subnet mask
255.255.255.0
Default gateway
192.168.1.212 or leave cells empty
1. Switch on the measuring device and connect to the computer via the cable .
2. If a 2nd network card is not used: all the applications on the notebook should be closed,
or all the applications that require the Internet or network, such as e-mail, SAP
applications, Internet or Windows Explorer, i.e. close all open Internet browsers.
3. Configure the properties of the Internet protocol (TCP/IP) as defined in the table above.