Technical data
Proline Promag W 400 HART
158
Hauser
16.8 Environment
Ambient temperature
range
Storage temperature
The storage temperature corresponds to the operating temperature range of the
measuring transmitter and the appropriate measuring sensors.→ 23
• Protect the measuring device against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
• Select a storage location where moisture cannot collect in the measuring device as
fungus or bacteria infestation can damage the liner.
• If protection caps or protective covers are mounted these should never be removed
before installing the measuring device.
Atmosphere
If a plastic transmitter housing is permanently exposed to certain steam and air mixtures,
this can damage the housing.
If you are unsure, please contact your Hauser Sales Center for clarification.
Degree of protection
Transmitter
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
Sensor
• As standard: IP66/67, type 4X enclosure
• Optionally available for remote version:
– IP66/67, type 4X enclosure; fully welded, with protective varnish EN ISO 12944 C5-M.
Suitable for use in corrosive atmospheres.
– IP68, type 6P enclosure; fully welded, with protective varnish as per EN ISO 12944 C5-
M. Suitable for permanent immersion in water ≤ 3 m (10 ft) or up to 48 hours at
depths ≤ 10 m (30 ft).
– IP68, type 6P enclosure; fully welded, with protective varnish as per EN ISO 12944
Im1/Im2/Im3. Suitable for permanent immersion in saline water ≤ 3 m (10 ft) or up to
48 hours at depths ≤ 10 m (30 ft) or in buried applications.
Vibration resistance
Compact version
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 2
000 Hz, 1 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.003 g
2
/Hz
– 200 to 2
000 Hz, 0.001 g
2
/Hz
– Total: 1.54 g rms
Remote version
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz, 7.5 mm peak
– 8.4 to 2
000 Hz, 2 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.01 g
2
/Hz
– 200 to 2
000 Hz, 0.003 g
2
/Hz
– Total: 1.54 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 50 g
Summary of Contents for Proline Promag W 400 HART