Diagnostics and troubleshooting
Proline Promag W 400 HART
122
Hauser
12
Diagnostics and troubleshooting
12.1 General troubleshooting
For local display
Error
Possible causes
Solution
Local display dark and no output
signals
Supply voltage does not match the
value indicated on the nameplate.
Apply the correct supply voltage
→ 45→ 45.
Local display dark and no output
signals
No contact between connecting
cables and terminals.
Check the connection of the cables
and correct if necessary.
Local display dark and no output
signals
Terminals are not plugged into the
main electronics module correctly.
Check terminals.
Local display dark and no output
signals
Main electronics module is
defective.
Local display dark and no output
signals
The connector between the main
electronics module and display
module is not plugged in correctly.
Check the connection and correct if
necessary.
Local display dark and no output
signals
The connecting cable is not plugged
in correctly.
1. Check the connection of the
electrode cable and correct if
necessary.
2. Check the connection of the coil
current cable and correct if
necessary.
Local display is dark, but signal
output is within the valid range
Display is set too bright or too dark. • Set the display brighter by
simultaneously pressing + .
• Set the display darker by
simultaneously pressing + .
Local display is dark, but signal
output is within the valid range
Display module is defective.
Backlighting of local display is red
Diagnostic event with "Alarm"
diagnostic behavior has occurred.
Take remedial measures
Text on local display appears in a
foreign language and cannot be
understood.
Incorrect operating language is
configured.
1. Press + for 2 s ("home
position").
2. Press .
3. Set the desired language in the
Display language
parameter
Message on local display:
"Communication Error"
"Check Electronics"
Communication between the
display module and the electronics
is interrupted.
• Check the cable and the
connector between the main
electronics module and display
module.
For output signals
Error
Possible causes
Solution
Signal output outside the valid
range
Main electronics module is
defective.
Device shows correct value on local
display, but signal output is
incorrect, though in the valid range.
Configuration error
Check and correct the parameter
configuration.
Device measures incorrectly.
Configuration error or device is
operated outside the application.
1. Check and correct parameter
configuration.
2. Observe limit values specified in
the "Technical Data".