Proline Promag L 400 HART
Commissioning
Hauser
89
ECC recovery time
The following option is
selected in the
Electrode cleaning
circuit
parameter:
On
Specify the recovery
time after electrode
cleaning to prevent
fluctuations in the
signal outputs.
The last flow
value measured
before the
cleaning is
retained.
1 to 600 s
5 s
ECC cleaning cycle
The following option is
selected in the
Electrode cleaning
circuit
parameter:
On
Specify the cleaning
cycle of the electrode
cleaning circuit.
30 to 10
080 min
40 min
ECC polarity
The following option is
selected in the
Electrode cleaning
circuit
parameter:
On
Select the polarity of
the electrode cleaning
circuit.
• Negative
• Positive
Negative (depending
on the electrode
material)
10.5.5 Configuring the totalizer
You can configure each totalizer in the three submenus
Totalizer 1-3
.
Navigation path
"Setup" menu
®
Advanced setup
®
Totalizer 1
"Setup" menu
®
Advanced setup
®
Totalizer 2
"Setup" menu
®
Advanced setup
®
Totalizer 3
Structure of the submenu
Totalizer 1-3
®
Assign process variable
Unit
Totalizer operation mode
Failure mode
Parameter overview with brief description
Parameters
Prerequisite
Description
Selection/
Factory setting
Assign process
variable
-
Select process
variable for totalizer.
Result
The selection
determines the
choose list of the
Unit
parameter.
• Off
• Volume flow
• Mass flow
Volume flow
Unit
One of the following options is
selected in the
Assign process
variable
parameter:
• Volume flow
• Mass flow
Select the unit for
the process variable
of the totalizer.
Unit choose list
Depends on country
and nominal
diameter