6 Block TOTALIZER
Device Functions Proline Promag 55 FOUNDATION Fieldbus
36
Hauser
6.1
Group TOTALIZER (1 to 3)
6.1.1
Function group CONFIGURATION
TOTALIZER
D
⇒
TOTALIZER 1
DAA
⇒
CONFIGURATION
300
⇓
TOTALIZER 2
DAB
⇒
CONFIGURATION
300
⇓
TOTALIZER 3
DAC
⇒
CONFIGURATION
300
Function description
TOTALIZER
→
TOTALIZER (1 to 3)
→
CONFIGURATION
The function descriptions below apply to Totalizers 1 to 3; the totalizers are independently configurable.
ASSIGN
(3000)
Use this function to assign a measured variable to the totalizer in question.
Options:
OFF
MASS FLOW
VOLUME FLOW
Factory setting:
VOLUME FLOW
!
Note!
• The totalizer is reset to "0" as soon as the selection is changed.
• If you select OFF, the function group CONFIGURATION (300) of the totalizer in
question only displays the ASSIGN (3000) function.
UNIT TOTALIZER
(3001)
Use this function to define the unit for the totalizer's measured variable, as selected
beforehand.
Options: (for MASS FLOW assignment):
Metric
→
g; kg; t
US
→
oz; lb; ton
Factory setting:
Nominal diameter and country dependent
Options (for VOLUME FLOW assignment):
Metric
→
cm
3
; dm
3
; m
3
; ml; l; hl; Ml Mega
US
→
cc; af; ft
3
; oz f; gal; Kgal; Mgal; bbl (normal fluids); bbl (beer);
bbl (petrochemicals); bbl (filling tanks)
Imperial
→
gal; Mgal; bbl (beer); bbl (petrochemicals)
Factory setting:
Nominal diameter and country dependent