5 Block USER INTERFACE
Device Functions Proline Promag 55 FOUNDATION Fieldbus
26
Hauser
5.2.2
Function group MULTIPLEX
USER INTERFACE
C
⇒
CONTROL
CAA
⇓
MAIN LINE
CCA
⇒
CONFIGURATION
220
⇓
MULTIPLEX
222
Function description
USER INTERFACE
→
MAIN LINE
→
MULTIPLEX
ASSIGN
(2220)
Use this function to define the second reading to be displayed in the main line alternately
(every 10 seconds) with the value defined in the ASSIGN function (2200).
Options:
OFF
VOLUME FLOW
MASS FLOW
VOLUME FLOW IN %
MASS FLOW IN %
TOTALIZER (1 to 3)
AI1 - OUT VALUE
AI2 - OUT VALUE
AI3 - OUT VALUE
AI4 - OUT VALUE
AI5 - OUT VALUE
PID - IN VALUE (controlled variable)
PID - CAS IN VALUE (external set point)
PID - OUT VALUE (manipulated variable)
Factory setting:
OFF
100%-VALUE
(2221)
!
Note!
This function is only available if VOLUME FLOW IN % or MASS FLOW IN % was
selected in the function ASSIGN (2220).
Use this function to define the flow value to be shown on the display as the 100% value.
User input:
5-digit floating-point number
Factory setting:
Nominal diameter and country dependent
FORMAT
(2222)
Use this function to define the maximum number of places after the decimal point for the
second value displayed in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
!
Note!
• Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device cannot always
be displayed, depending on this setting and the engineering unit. In these instances an
arrow appears on the display between the measured value and the engineering unit
(e.g. 1.2
→
kg/h), indicating that the measuring system is computing with more
decimal places than can be shown on the display.