Wiring
Promag 50
48
Hauser
Cable termination for the remote version
Promag H
Terminate the signal and coil current cables as shown in the figure below (Detail A).
Ferrules must be provided on the fine-wire cores (Detail B:
m
= red ferrules,
1.0 mm;
n
= white ferrules,
0.5 mm).
"
Caution!
When fitting the connectors, pay attention to the following points:
•
Signal cable
Make sure that the ferrules do not touch the wire shield on the sensor side.
Minimum distance = 1 mm (exception "GND" = green cable).
•
Coil current cable
Insulate one core of the three-core wire at the level of the core reinforcement; you only require two cores for the connection.
• On the sensor side, reverse both cable shields approx. 15 mm over the outer jacket. The strain relief ensures an electrical connection with the connection
housing.
TRANSMITTER
Signal cable
Coil current cable
A0002686
A0002684
SENSOR
Signal cable
Coil current cable
A0002647
A0002648
A
B
GND
m
m
m
m
n
n
n
n
80 (3.15)
50 (1.97)
17 (0.67)
8 (0.31)
mm (inch)
A
B
m
m
m
70 (2.76)
50 (1.97)
8 (0.31)
mm (inch)
10 (0.39)
A
B
GND
m
n
n
n
mm (inch)
80 (3.15)
15 (0.59)
8 (0.31)
17 (0.67)
³
1 (0.04)
³
A
B
m
m
70 (2.76)
15 (
)
0.59
8 (0.31)
40 (1.57)
mm (inch)