Promag 50
Technical data
Hauser
99
Adapters
Length of connecting cable
10.1.8
Operating conditions: Environment
Ambient temperature range
• Transmitter:
– Standard: –20 to +60 °C (–4 to +140 °F)
– Optional: –40 to +60 °C (–40 to +140 °F)
!
Note!
At ambient temperatures below –20 (–4 °F) the readability of the display may be impaired.
• Sensor (the ambient temperatur range is dependent on the sensor type):
– Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
– Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)
"
Caution!
• The permitted temperature range of the measuring tube lining may not be undershot or overshot
(
"Operating conditions: Process"
"Medium temperature range").
• Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• The transmitter must be mounted separate from the sensor if both the ambient and fluid
temperatures are high.
Storage temperature
The storage temperature corresponds to the operating temperature range of the measuring
transmitter and the appropriate measuring sensors.
"
Caution!
• The measuring device must be protected against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
• A storage location must be selected where moisture does not collect in the measuring device. This
will help prevent fungus and bacteria infestation which can damage the liner.
Degree of protection
• Standard: IP 67 (NEMA 4X) for transmitter and sensor
• Optional: IP 68 (NEMA 6P) for remote version of Promag L, W and P sensor.
Promag L only with stainless steel flanges.
Shock and vibration resistance
Acceleration up to 2 g following IEC 60068-2-6
(high-temperature version: no data available)
CIP cleaning
"
Caution!
The maximum fluid temperature permitted for the device may not be exceeded.
CIP cleaning is possible:
Promag P, Promag H
CIP cleaning is not possible:
Promag D, Promag L, Promag W
SIP cleaning
"
Caution!
The maximum fluid temperature permitted for the device may not be exceeded.
SIP cleaning is possible:
Promag H
SIP cleaning is not possible:
Promag D, Promag L, Promag W, Promag P