Promag 50
Technical data
Hauser
101
a0002671
Abb. 59:
Remote version Promag P (with PFA- or PTFE-lining)
T
A
= ambient temperature; T
F
= fluid temperature; HT = high-temperature version with insulation
m
= light gray area
temperature range from –10 to –40 °C (–14 to –40 °F) is valid for stainless steel version only
n
= diagonal hatched area
foam lining (HE) and degree of protection IP68 = fluid temperature
max. 130°C / 266 °F
Promag H
Sensor:
• DN 2 to 25: –20 to +150 °C (–4 to +302 °F)
• DN 40 to 100: –20 to +150 °C (–4 to +302 °F)
Seals:
• EPDM: –20 to +150 °C (–4 to +302 °F)
• Silicone: –20 to +150 °C (–4 to +302 °F)
• Viton: –20 to +150 °C (–4 to +302 °F)
• Kalrez: –20 to +150 °C (–4 to +302 °F)
Conductivity
The minimum conductivity is
5
μ
S/cm (
20
μ
S/cm for demineralized water)
!
Note!
Note that in the case of the remote version, the requisite minimum conductivity is also influenced
by the length of the connecting cable
Medium pressure range
(nominal pressure)
Promag D
• EN 1092-1 (DIN 2501)
– PN 16
• ANSI B 16.5
– Class 150
• JIS B2220
– 10 K
Promag L
• EN 1092-1 (DIN 2501)
– PN 6 (DN 350…1200)
– PN 10 (DN 50…1200)
– PN 16 (DN 50…150)
• EN 1092-1, lap joint flange, stampel plate
– PN 10 (DN 50 to 300)
HT
PTFE
m
n
PFA
0
0
0
0
-20
-40
-40
20
20
40
100
60
140
[°F]
[°C]
TA
40
60
80
TF
100
100
120 140 160
200
300
180
[°C]
360 [°F]
-20
-40
-40