Description of device parameters
Proline Promag 300 Modbus RS485
92
Hauser
Additional information
Description
Corrected value = (factor × value) + offset
Conduct. factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Conduct. factor (1849)
Prerequisite
The
On
option is selected in the
Conduct. measur.
parameter (→ 74) parameter.
Description
Use this function to enter a quantity factor for the conductivity. This multiplication factor
is applied over the conductivity range.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset
Corr. vol offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol offset (1866)
Description
Use this function to enter the zero point shift for the corrected volume flow trim. The
corrected volume flow unit on which the shift is based is 1 Nm³/s.
User entry
Signed floating-point number
Factory setting
0 Nm³/s
Additional information
Description
Corrected value = (factor × value) + offset
Corr. vol factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol factor (1867)
Description
Use this function to enter a quantity factor (without time) for the corrected volume flow.
This multiplication factor is applied over the corrected volume flow range.
User entry
Positive floating-point number
Factory setting
1