Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
89
Install. direct.
Navigation
Expert → Sensor → Sensor adjustm. → Install. direct. (1809)
Description
Use this function to change the sign of the medium flow direction.
Selection
• In arrow direct.
• Against arrow
Factory setting
In arrow direct.
Additional information
Description
Before changing the sign: ascertain the actual direction of fluid flow with reference to
the direction indicated by the arrow on the sensor nameplate.
Integration time
Navigation
Expert → Sensor → Sensor adjustm. → Integration time (6533)
Description
Display the duration of an integration cycle.
User interface
1 to 65 ms
Measuring period
Navigation
Expert → Sensor → Sensor adjustm. → Measuring period (6536)
Description
Display the time of a full measuring period.
User interface
0 to 1
000 ms
"Variable adjust" submenu
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust
‣
Variable adjust
Vol. flow offset (1831)
Vol. flow factor (1832)
Mass flow offset (1841)