Description of device parameters
Proline Promag 300 Modbus RS485
86
Hauser
Additional information
Dependency
The unit is taken from the
Density unit
Ref. temperature
Navigation
Expert → Sensor → External comp. → Ref. temperature (1816)
Prerequisite
The
Fixed density
option or
External density
option are selected in the
Density source
parameter (→ 85).
Description
Use this function to enter a reference temperature for calculating the reference density.
User interface
–273.15 to 99
999 °C
Factory setting
Country-specific:
• +20 °C
• +68 °F
Additional information
Dependency
The unit is taken from the
Temperature unit
Reference density calculation
ρ
n
ρ
α
β
(1 +
+
⋅ ∆
t
⋅
⋅ ∆
t
)
2
=
A0023403
• ρ
N
: reference density
• ρ: fluid density currently measured
• t: fluid temperature currently measured
• t
N
: reference temperature at which the reference density is calculated (e.g. 20 °C)
• ∆t: t - t
N
• α: linear expansion coefficient of the fluid, unit = [1/K]; K = Kelvin
• β: square expansion coefficient of the fluid, unit = [1/K
2
]
Ref. density
Navigation
Expert → Sensor → External comp. → Ref. density (1892)
Prerequisite
The
Calculated value
option is selected in the
Density source
parameter.
Description
Displays the reference density.
User interface
Positive floating-point number