Oxymax COS61D
Maintenance
Hauser
27
Clean the sensor:
• Before every calibration
• At regular intervals during operation as necessary
• Before returning it for repairs
Type of contamination
Cleaning
Salt deposits
1.
Immerse the sensor in drinking water or in 1-5 % hydrochloric
acid (for a few minutes).
2.
Then rinse it with copious amounts of water.
Dirt particles on the sensor shaft and
shaft sleeve (
not fluorescence cap!
)
‣
Clean sensor shaft and sleeve with water and a suitable sponge.
Dirt particles on fluoresence cap
‣
Clean the fluorescence cap with water and a soft cloth.
‣
After cleaning:
Rinse with copious amounts of clean water.
10.4 Cleaning sensor optics
The optics need to be cleaned only if medium has penetrated a defective fluorescence cap.
1. Unscrew the protection guard and fluorescence cap from the sensor head.
2. Carefully clean the optical surface with a soft cloth until the buildup is fully removed.
3. Wipe the optical surface with a soft cloth that is wetted with drinking water or
distilled water.
4. Dry the optical surface and screw on a functional fluorescence cap.
5. At the transmitter, execute the
Sensor cap change
command and then perform the
necessary calibrations.
NOTICE
Damage, scratches on optical surface
Distorted measured values
‣
Make sure that the optical surface is not scratched or damaged in any other way.
10.5 Wear parts and consumables
Parts of the sensor are subject to wear during operation. By taking suitable measures, you
can restore the normal operating function.
Corrective action
Reason
Replace process seals
Visible damage to a process seal
Replace fluorescence cap Fluorescence layer is damaged or can no longer be cleaned (black protective layer is
damaged to the point that the pink fluorescence layer is visible)
10.5.1 Replacing sealing rings
It is compulsory to replace the sealing ring if it is visibly damaged. Only use original sealing
rings (O-ring kit: Order No.: 51518597).
10.5.2 Replacing the fluorescence cap
The typical operating life of a fluorescence cap is more than 2 years. The sensor checks
whether the cap is aging and issues a warning via the transmitter if the rate of aging
Summary of Contents for Oxymax COS61D
Page 36: ...Technical data Oxymax COS61D 36 Endress Hauser Interface Memosens protocol...
Page 37: ...Oxymax COS61D Appendices Endress Hauser 37 14 Appendices...
Page 38: ...Appendices Oxymax COS61D 38 Endress Hauser A0044703...
Page 41: ......
Page 42: ......
Page 43: ......