5.3
Post-installation check
Put the sensor into operation only if you can answer "yes" to the following questions:
• Are the sensor and cable undamaged?
• Have you chosen a correct mounting angle?
6
Electrical connection
L
WARNING
Device is live
Incorrect connection may result in injury or death
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣
Prior
to commencing connection work, ensure that no voltage is present on any cable.
6.1
Connecting the sensor
The sensor is connected to the transmitter using the pre-terminated or labeled sensor fixed
cable.
YE
RD
WH
YE
BK
BK
GY
A0029261
8
OUSAF11 fixed cable
CM44P terminal
CVM40 terminal
Cable color
Assignment
P+
V1.1
YE (thick)
Lamp v
S+
V1.3
YE (thin)
Detection of lamp v
S-
V1.4
BK (thin)
Detection of lamp voltage -
P-
V1.2
BK (thick)
Lamp voltage -
A (1)
S1.1
RD
C(1)
S1.2
WH
Sensor -
SH (1)
S1.S
GY
Shield