Technical data
LPGmass Modbus RS485
76
Hauser
Function codes
• 03: Read holding register
• 04: Read input register
• 06: Write single registers
• 08: Diagnostics
• 16: Write multiple registers
• 23: Read/write multiple registers
Broadcast messages
Supported by the following function codes:
• 06: Write single registers
• 16: Write multiple registers
• 23: Read/write multiple registers
Supported baud rate
• 1
200 BAUD
• 2
400 BAUD
• 4
800 BAUD
• 9
600 BAUD
• 19
200 BAUD
• 38
400 BAUD
• 57
600 BAUD
• 115
200 BAUD
Data transfer mode
• ASCII
• RTU
Data access
Each device parameter can be accessed via Modbus RS485.
For Modbus register information, see "Description of device parameters"
documentation
16.5 Power supply
Terminal assignment
Supply voltage
The power unit must be tested to ensure it meets safety requirements (e.g. PELV, SELV).
Transmitter
• Modbus RS485, for use in non-hazardous areas and Zone 2/Div. 2:
DC 20 to 30 V
• Modbus RS485, for use in intrinsically safe areas:
Power supply via Safety Barrier Promass 100
Safety Barrier Promass 100
DC 20 to 30 V
Power consumption
Transmitter
Order code for "Output"
Maximum
Power consumption
Option
M
: Modbus RS485, for use in intrinsically safe areas
2.45 W
Safety Barrier Promass 100
Order code for "Output"
Maximum
Power consumption
Option
M
: Modbus RS485, for use in intrinsically safe areas
4.8 W