MMP20 SONO-DIS (option D)
Hauser
Page 21
8.5.5
Display contrast
At extreme temperatures, it may be necessary to change the contrast setting to improve screen
readability.
After selecting the "Display contrast" menu item, change the contrast setting for the display with the
"Up"
or "Down"
key. Set the contrast so that you can clearly see all the gray gradations in the
bar chart. Press the "Measure" key
to activate and save your settings. Once the setting has been
activated, the symbol appears in the top right-hand corner of the display.
8.5.6
Probe info
Information about the probe is displayed shortly after selecting this menu item.
This information includes:
•
Serial number
•
Probe type
•
Hardware revision (HW)
•
Firmware revision (FW)
8.5.7
SONO-DIS info
Information about your SONO-DIS handheld device is displayed if this menu item is selected.
This information includes:
•
Serial number
•
Hardware revision (HW)
•
Firmware revision (FW)
•
Battery capacity
•
Battery voltage
8.5.8
Material calibration curves
Selecting this menu item allows you to set the SONO-WZ probe to another material calibration
curve. When the SONO-DIS is switched on, the calibration curve set in this menu item is displayed
at the bottom of the screen for approx. 3 seconds.
Caution:
The standard calibration curve
"Cal. No.: 4"
is set by default for concrete. This setting
should not be changed, or should only be changed if you are measuring a material other than fresh
concrete.
Up to 15 calibration curves in total can be managed for materials such as suspensions, sludges etc.
Furthermore, the sensitivity of the concrete measurement can be changed by selecting another
calibration curve. Please contact the Service Department of the manufacturer for more details.