Mounting
Micropilot M FMR244 with FOUNDATION Fieldbus
18
Hauser
3.4
Installation instructions
3.4.1
Mounting kit
Beside a key AF 60 for installation you need the following tool:
• 4 mm/0.1" Allen wrench for turning the housing.
3.4.2
Installation in tank (free space)
Optimum mounting position
L00-FMR244xx-17-00-00-en-003
90°
90
°
90°
90°
90°
90
°
90°
90°
9
90°
90°
60
mark at threaded boss
G 1½”
or
1½ NPT
AF 60
max. torque
20 Nm
Caution!
Use only the
threaded boss
Standard installation
When mounting in a tank, please observe
engineering hints on Page 13 and the following
points:
• Marker is aligned towards tank wall.
• Install the device using the threaded boss (AF
60) only. Observe the max. torque of 20 Nm.
• After mounting, the housing can be turned
350° in order to simplify access to the display
and the terminal compartment.
• For optimum measurement, the horn antenna
should extend below the nozzle. Nozzle
heights up to 500 mm can be accepted if this
should not be possible due to mechanical
reasons.
Note!
Please contact Hauser for application
with higher nozzle.
• The antenna must be aligned vertically.
L00-FMR244xx-17-00-00-de-002
H
Ø D
Antenna size
1½" / 40 mm
D [mm/inch]
39 / 1.5
H [mm/inch]
< 85 / < 3.4