12.3 Diagnostic event in the operating tool
If a diagnostic event is present in the device, the status signal appears in the top left status
in the operating tool along with the corresponding symbol for event level in accordance
with NAMUR NE 107:
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
Calling up remedial measures
1. Navigate to the
Diagnostics
menu.
In the
Actual diagnostics
parameter, the diagnostic event is shown with event
text.
2. On the right in the display range, hover the cursor over the
Actual diagnostics
parameter.
A tool tip with remedial measures for the diagnostic event appears.
12.4 Diagnostic list
In the Diagnostic list submenu, up to 5 currently pending diagnostic messages can be
displayed. If more than 5 messages are pending, the messages with the highest priority are
shown on the display.
Navigation path
Diagnostics → Diagnostic list
Calling up and closing the remedial measures
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message about the remedial measures closes.