Distance for liquid measurements
The unit is defined in the
Distance unit
Signal quality
Navigation
Setup → Signal quality
Description
Displays the signal quality of the level echo.
Additional information
Meaning of the display options
•
Strong
The evaluated echo exceeds the threshold by at least 10 dB.
•
Medium
The evaluated echo exceeds the threshold by at least 5 dB.
•
Weak
The evaluated echo exceeds the threshold by less than 5 dB.
•
No signal
The device does not find a usable echo.
The signal quality indicated in this parameter always refers to the currently evaluated
echo: either the level echo or the tank bottome echo. To differentiate between these two,
the quality of the tank bottom echo echo is always displayed in brackets.
In case of a lost echo (
Signal quality = No signal
) the device generates the following
error message:
• F941, for
Output echo lost (→ 134) = Alarm
.
• S941, if another option has been selected in
.
Confirm distance
Navigation
Setup → Confirm distance
Description
Specify, whether the measured distance matches the real distance.
Depending on the selection the device automatically sets the range of mapping.