Electrical connection
Memosens COS81E
20
Hauser
6
Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣
Prior
to commencing connection work, ensure that no voltage is present on any cable.
6.1
Connecting the sensor
The electrical connection of the sensor to the transmitter is established using measuring
cable CYK10.
GN/YE
YE
GN
BN
WH
GND
+
–
Com A
Com B
GY
A0024019
12
Measuring cable CYK10
6.2
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions
and which are necessary for the required, designated use, may be carried out on the device
delivered.
‣
Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due,
for example to covers being left off or cable (ends) that are loose or insufficiently secured.
6.3
Post-connection check
Device condition and specifications
Action
Are the sensor, assembly or cables free from damage on
the outside?
‣
Perform a visual inspection.
Electrical connection
Action
Are the mounted cables strain-relieved and not twisted?
‣
Perform a visual inspection.
‣
Untwist the cables.
Is a sufficient length of the cable cores stripped, and are
the cores positioned in the terminal correctly?
‣
Perform a visual inspection.
‣
Pull gently to check they are seated correctly.
Are all the screw terminals properly tightened?
‣
Tighten the screw terminals.
Are all cable entries mounted, tightened and leak-tight? In the case of lateral cable entries:
‣
Point cable loops downward so that water can
drip off.
Are all cable entries installed downwards or mounted
laterally?