Repair
Memosens COS51E
38
Hauser
3. Confirm and save the process.
The sensor is ready for operation.
After replacing the sensor cap:
1. Re-immerse the sensor into the medium.
2. Reset the counter on the transmitter.
e.g. CM44x:
Menu
/
Calibration
/
Oxygen (amp.)
/
Sensor cap change
3. Confirm and save the process.
The sensor is ready for operation.
10.4 Checking the measurement function
1. Remove the sensor from the medium.
2. Clean and dry the membrane.
3. Adjust the process pressure at the transmitter if it differs from the atmospheric
pressure; otherwise, a comparison will not be possible.
4. After about 10 minutes, measure the oxygen saturation index in air (without
recalibration).
The measured value should be at 100 ± 2 % SAT.
10.5 Disposal
If required by the Directive 2012/19/EU on waste electrical and electronic equipment
(WEEE), the product is marked with the depicted symbol in order to minimize the
disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this
marking as unsorted municipal waste. Instead, return them to Hauser for
disposal under the applicable conditions.