Memosens COS51E
Diagnostics and troubleshooting
Hauser
29
Problem
Test
Remedial action
Is membrane coated?
‣
Clean sensor carefully.
Display value fluctuating
Membrane visibly stretched?
‣
Replace membrane cap.
Follow the troubleshooting instructions in the Operating Instructions for the
transmitter. Check the transmitter if necessary.
If one of the above mentioned faults occurs:
1. Unscrew membrane cap from sensor.
2. Dry the electrodes.
Does the transmitter display 0?
If the transmitter does
not
display 0:
3. Check the electrical connection.
Does the transmitter display 0?
If the transmitter does
not
display 0:
4. Contact the Hauser Service department.