Mounting
Memosens COS51E
18
Hauser
Basin rim mounting with immersion tube
1
2
3
4
5
A0042860
11
Basin rim mounting
1
Pendulum holder CYH112
2
Assembly Flexdip CYA112
3
Assembly float
4
Oxygen sensor
5
Basin rim
Float
The CYA112 float is for use in the case of large fluctuations in water level, for example in
rivers or lakes.
1
2
3
4
5
6
7
G1
800 (31.50)
500 (19.69)
220 (8.66)
Ø200 (7.87)
A0032159
12
Dimensions in mm (inch)
1
Cable run with strain relief and rain shield
2
Fixing ring for rope and chains with terminal screw
3
Eyelets Ø15, 3 x 120 ° for anchoring
4
Plastic float, resistant to salt water
5
Pipe 40 x 1, stainless steel 1.4571
6
Bumper and ballast
7
Oxygen sensor
5.3.3
Flow assembly COA250
The COA250 flow assembly with automatic self-venting is suitable for use in pipelines or
at hose connections. The inlet is at the bottom of the assembly and the outlet is at the top
(G¾ connection thread). It is installed in a pipe using two 90° pipe brackets at the inlet of
assembly (item 6).